Troubleshooting
4
6000 X-Series Oscilloscopes Service Guide
77
To check power to the acquisition board
This procedure checks the power supplies from the power board to the acquisition
board. Values outside the expected range help identify bad assemblies.
1
“To prepare for internal assembly
2
Connect the negative lead of the multimeter to a ground point on the
oscilloscope.
3
Connect the power cord, and turn on the oscilloscope.
4
First check the bulk power supply voltage:
5
Next, check the supplies from the power board to the acquisition board:
Figure 15
Location of the J105 acquisition board power connector
J105
1
10
20
11