Testing Performance
2
6000 X-Series Oscilloscopes Service Guide
31
3
Add a measurement for the average voltage:
a
Press the
[Meas]
key.
b
Press
Source
; then, turn the Entry knob (labeled
on the front panel) to
select the channel you are testing.
c
Press
Type:
; then, turn the Entry knob to select
Average - Full Screen
, and press
Add Measurement
.
4
Read the “current” average voltage value as V1.
5
Use the BNC tee and cables to connect the precision source / power supply to
both the oscilloscope and the multimeter (see
Note: Set the Low Force terminal of the Precision Source to its "Floating" state
to prevent offset error caused by ground loop current from the Precision Source
ground to the DUT ground.