26
6000 X-Series Oscilloscopes Service Guide
2
Testing Performance
Figure 2
Setting Up Equipment for Digital Channel Threshold Accuracy Test
4
Use a BNC-banana cable to connect the multimeter to the other side of the
BNC Tee.
5
Connect the BNC Tee to the Channel 1 output of the precision source as shown
.
6
On the oscilloscope, press the
[Digital]
key, then press the
Thresholds
softkey,
then press the
D7 - D0
softkey repeatedly until the check mark is next to
User
.
Probe
Cables
Channels
8 - 15
Channels
0 - 7
Test
Connector
6000 X-Series Oscilloscope
Precision Source
Digital
Multimeter
BNC Tee
BNC-Bananna