16 Keysight E8363B/C and E8364B/C Installation Note
General Information
Installation Procedure for the Upgrade Kit
1-
4.
Attach one step attenuator to each bracket using two M3.0 x 8 screws
(provided) for each.
Be careful to position the step attenuators so that the necessary cables
can be attached. The end of the step attenuator with the ribbon cable
connector must face toward the inside of the analyzer. Review the cable
connections in
if necessary.
5.
Using a T-8 TORX driver, attach one bias tee to each bracket using two
M2.5 x 14 screws (provided) for each.
The bias tees mount beneath the detectors. Be careful to position the bias
tees so that the necessary cables can be attached. The end of the bias tee
with the wires attached must face toward the inside of the analyzer.
Review the cable connections in
6.
Reinstall the brackets in the analyzer with the step attenuators, bias tees,
and detectors attached.
7.
Reconnect the stranded control cables, (item
①
), and the flexible RF
cables, (item
✍
) to the detectors.
Figure 9
Attenuator and Bias Tee Installation, Bracket Installation