28 Keysight E8363B/C and E8364B/C Installation Note
General Information
Installation Procedure for the Upgrade Kit
1-
Step 12.Perform Post-Upgrade Adjustments and Calibration
The following adjustments must be made due to the change in the full
frequency range of the analyzer.
— source calibration
— receiver calibration
These adjustments are described in the PNA service guide and in the PNA
on-line HELP. A list of equipment required to perform these adjustments can
be found at
“Equipment Required for Post-Upgrade Adjustments” on page 6
.
Performance Tests and System Verification
The analyzer should now operate and phase lock over its entire frequency
range.
If you experience difficulty with the basic functioning of the analyzer, contact
Keysight. Refer to
“Getting Assistance from Keysight” on page 1
Although the analyzer functions, its performance relative to its specifications
has not been verified for the additional frequency range enabled by this
upgrade.
It is recommended that a full instrument calibration be performed using the
PNA performance test software.
If the testing of the analyzer’s full range of specifications is not required, a
system verification can be performed.
Refer to the analyzer’s service guide for information on performance tests and
system verification.