Keysight16197A Operation and Service Manual 71
Service
Operation Check
Step 4.
Obtain the difference between the first and second Ls measurements.
Step 5.
Confirm that the variation in measurements is within the typical range shown
in
Table 6-8
.
Table 6-8
SHORT Impedance Check: Typical
Parameter
Frequency
Typical (Difference)
Ls
100MHz
± 45pH
Ls
1GHz
± 45pH