Keysight16197A Operation and Service Manual 27
Operation
Changing the Orientation of the Device Guide and Electrode Plate
Figure 3-3
Dimension of the Device Guide Frame
Be sure to superpose the device guide onto the electrode plate in the
proper orientation and with the correct side facing up. Otherwise,
adequate electrode spacing cannot be created. As shown in
Figure 3-3
when viewed from above, on the correct face of the device guide there are
two marks at positions 90 degrees apart clockwise from a single position.
Superpose the device guide onto the electrode plate so that all these
marks are aligned with the corresponding marks on the electrode plate.
For a 0603 device, correct electrode spacing can be created regardless of
whether the mark on the device guide is aligned with the corresponding
mark on the electrode plate