Keysight16197A Operation and Service Manual 51
User Maintenance
Check Sheet
Check Sheet
Example Entry in Check Sheet
The following tables shown an example entry of electrode deterioration check
results in the Check Sheet.
Example Entry of Electrode Deterioration Check Results
DUT Size
2012
Table 4-2
Example Entry of Reference Values and User Limit Values
Frequency
a
Measurement
parameter
Reference
value [a]
b
User limit
value [b]
c
Lower limit
[a-b]
Upper limit
[a+b]
100 MHz
Rs
40 m
30 m
10 m
70 m
Ls
2.34 nH
0.5 nH
1.84 nH
2.84 nH
800 MHz
Rs
220 m
40 m
180 m
260 m
Ls
2.32 nH
0.4 nH
1.92 nH
2.72 nH
a. User's discretionary value
b. Enter the values obtained in section
“Acquiring Reference Values” on
page 48
c. For entry, see
“Example Settings of User Limit Values” on page 47
Table 4-3
Example Entry of Checking History
Date/
Time
Frequency
Measurement
parameter
Measurement
value
Acceptability
Set position
of pressure
adjusting
nut
2000/
10/11
9:30
100 MHz
Rs
50 mW
O
Ls
2.5 nH
O
2000/
10/11
9:35
800 MHz
Rs
250 mW
O
Ls
2.6 nH
O
2000/
10/12
9:30
100 MHz
Rs
55 mW
O
Ls
2.6 nH
O
2000/
10/12
9:35
800 MHz
Rs
285 mW
X
a
Ls
2.7 nH
O
a.
Replace parts if check results are not acceptable.
Содержание 16197A
Страница 1: ...Operation and Service Manual Keysight 16197A Bottom Electrode SMD Test Fixture...
Страница 10: ...10 Keysight16197A Operation and Service Manual Installation Guide Storage 1 Figure 1 5 Pressure Adjusting Nut...
Страница 22: ...22 Keysight16197A Operation and Service Manual Product Description Product Description 2...
Страница 42: ...42 Keysight16197A Operation and Service Manual Operation Connecting and Measuring the DUT 3...
Страница 54: ...54 Keysight16197A Operation and Service Manual User Maintenance Assembly Check 4...
Страница 72: ...72 Keysight16197A Operation and Service Manual Service Operation Check 6...