Keysight16197A Operation and Service Manual 37
Operation
Performing Fixture Compensation
Step 5.
Press the lever; the latch is disengaged (
Figure 3-16,1
). Slowly release the
lever to lower the pressure rod (
Figure 3-16,2
).
Figure 3-16
Setting the Fixture for SHORT State Using a Shorting Bar
Step 6.
Measure the SHORT compensation data according to the operation manual
for the measuring instrument to be used.
Shorting bars can wear out. Therefore, each time before using a shorting
bar, measure and compare its resistance with that of a new shorting bar.
For details, see
“Operation Check” on page 69
Adjust the position of the pressure unit in advance so that a shorting bar
can be pressed at its center by the end of the pressure rod.
When measurement is not being done, lower the pressure rod so it rests on
an unoccupied area of the device guide.