Keysight16197A Operation and Service Manual 69
Service
Operation Check
Operation Check
This section describes the operation check method. Be sure to perform the
operation check after each time a part is replaced.
Open Impedance Check
Conduct an Open impedance check with the fixture set for the OPEN state.
Necessary
equipment
— Shorting bars (supplied as accessories)
— Impedance measuring instrument (calibrated at the 7-mm connector end)
Step 1.
Prepare a measuring instrument calibrated at the 7-mm connector end.
Connect the test fixture with this instrument.
Step 2.
Set the fixture for the OPEN state, at the DUT measuring position (see
Figure
3-18 on page 38
).
Step 3.
Make the settings for the 4291B as follows.
Step 4.
Under these conditions, measure Cp at 100 MHz and 1 GHz separately.
Step 5.
Confirm that the Cp value is within the typical range shown in Table 6-5.
If a measuring instrument other than the 4291B is to be used, make the equivalent
settings according to the Operation and Service Manual for the instrument to be
used.
Table 6-4
Settings for the Measuring Instrument (Keysight 4291B)
Measuring Condition
Setting
Measurement parameter
Cp
Start frequency
100MHz
Stop frequency
1GHz
OSC Level
0.5V
Number of points
2
Point averaging factor
16
Points averaging
ON
Table 6-5
OPEN Impedance Check: Typical
Parameter
Frequency
Typical (Absolute value)
Cp
100MHz
700 fF ± 400 fF
Cp
1GHz
700 fF ± 400 fF
Содержание 16197A
Страница 1: ...Operation and Service Manual Keysight 16197A Bottom Electrode SMD Test Fixture...
Страница 10: ...10 Keysight16197A Operation and Service Manual Installation Guide Storage 1 Figure 1 5 Pressure Adjusting Nut...
Страница 22: ...22 Keysight16197A Operation and Service Manual Product Description Product Description 2...
Страница 42: ...42 Keysight16197A Operation and Service Manual Operation Connecting and Measuring the DUT 3...
Страница 54: ...54 Keysight16197A Operation and Service Manual User Maintenance Assembly Check 4...
Страница 72: ...72 Keysight16197A Operation and Service Manual Service Operation Check 6...