58 Keysight16197A Operation and Service Manual
Specifications and Supplemental Performance Characteristics
Rs (ESR) Measurement
5-
Rs (ESR) Measurement
Additional error Rse[%] of the Rs measurement is calculated by using the
additional error for Impedance Ze [%] as follows.
If Dx
0.1:
Rse [%] = Ze / Dx
If 0.1 < Dx
0.5:
Rse [%] = (Ze / Dx) ×
Dx is the measured value of D and is calculated as follows.
Dx = 2 ×
π
× f × Csx × Rsx,
where
f: measurement signal frequency
Csx: measured value of Cs
Rsx: measured value of Rs
Figure 5-1 Additional Error for Impedance
1 Dx
2
+