24
Mains Fail (Optional extra feature)
•
When the input configured as Mains Fail is activated
•
Screen Backlight Red
•
Mains Fail is displayed
•
Audible alarm sounds
•
Audible can be muted via Enter pushbutton -- this silences the alarm if configured.
Note - Not for use with Relay Interface unit -- AFA1000/E only.
Fan Stop
•
When the input configured as Fan Stop is activated
•
Off is displayed
•
Screen Backlight Red
•
Audible alarm muted
•
When the input configured as Fan Stop is de-activated
•
Start Up Timer is displayed
•
Audible alarm and relays inhibited until Start up time has elapsed
Note - Used if the AFA1000 is a Slave unit with a common blower (Remote master On/Off signal is
used)
Mute
•
When the input configured as Mute and is activated
•
Audible alarm muted
•
Mute Icon shown on display
Note - The Mute function disables the pushbutton mute so that any audible alarm can only be muted
using the input e.g. remote key switch.