2.9 External Connections
The airflow monitor has the following Inputs:-
Input 1
Volt free input configurable for normally closed, normally open relays or Analogue 0-5VDC Input
This input can be configured as:-
Digital Input Functions (Closed or Open volt free contact):-
Alarm disable
Night set-back
External alarm
Emergency
Sash High
High / Low
Sash Warning
Mains Fail
Fan Stop
Mute
Analogue Input Functions:-
Temperature
Input 2
Volt free input configurable for normally closed, normally open relays or Analogue 0-5VDC Input
This input can be configured as:-
Digital Input Functions (Closed or Open volt free contact):-
Alarm disable
Night set-back
External alarm
Emergency
Sash High
High / Low
Sash Warning
Mains Fail
Fan Stop
Mute
Analogue Input Functions:-
Temperature