23
External alarm
•
When input configured as External alarm is activated
•
Screen Backlight Red
•
External Alarm toggles on /off with display -- (if configured)
•
Audible alarm sounds – can be muted via Enter pushbutton
•
External alarm relay operates (if configured)
Emergency
•
When input configured as Emergency is activated
•
Screen Backlight Red
•
Emergency toggles on /off with display -- (if configured)
•
Audible alarm sounds – can be muted via Enter pushbutton
•
Emergency alarm relay operates ( if configured)
Sash High
•
When input configured as Sash High is activated
•
Screen Backlight Amber
•
Sash High toggles on / off with display if sash height > calibrated position
•
Audible alarm sounds -- can be muted via Enter pushbutton
•
Sash High relay operates (if configured)
•
Alarm re-activates after time delay if repeat time enabled
High / Low
•
When input configured as High/Low is activated
•
Display Icon shows Up (High) or Down (Low) arrows
•
High / Low relay operates ( if configured )
This function is designed for two speed fan operation or two position damper operation switched via a
micro switch or proximity switch activated at a given position on the sash.
Close Sash
•
When the input configured as Close Sash is activated
•
Screen Backlight Red
•
Close Sash – toggles on / off with velocity display
•
Audible alarm sounds (after pre-set time)
•
Audible can be muted via Enter pushbutton -- this silences the alarm if configured.
Reset when Sash lowered to closed position and input de-activated or operator is present
Содержание Air Alert 600
Страница 2: ...2 Issue 02 June 2017...
Страница 5: ...5 1 SAFETY...
Страница 10: ...10 2 Overview of the Airflow Monitor...
Страница 20: ...20 3 Functions Operation...
Страница 25: ...25 4 Components...
Страница 28: ...28 5 Installation...
Страница 33: ...33 6 Configuration Calibration...
Страница 37: ...37 7 Auxiliary Features Connections...
Страница 40: ...40 8 RS485 Coms Output...
Страница 47: ...47 9 Warranty...