2.3 Displayed Alarms and Events
Sash High – will be displayed when the Sash alarm is enabled and the sash is raised above
the max safe working opening. Sash High will alternate on/off with the velocity
reading. The Screen color will change to Amber.
Ext Alarm –
will be displayed when the external alarm input is activated (when enabled)
Ext Alarm will alternate on/off with the velocity reading. The Screen color will change
to Red.
Air Fail -
will be displayed if the airflow is less than the Low air alarm point.
Air Fail will alternate on/off with the velocity reading. The Screen color will change
to Red.
High Air -
will be displayed if the airflow is more than the High air alarm point.
High Air will alternate on/off with the velocity reading. The Screen color will change
to Red.
Set-back- will be displayed if the night set-back function is activated. Setback will alternate
on/off with the velocity reading and the Screen color will change to Red and the
backlight will dim.
Alm Dis-
will be displayed if the alarm disable function is activated (when enabled)
Alarm Disable will alternate on/off with the velocity reading. The Screen color will
change to Red.
Close Sash -
will be displayed if the sash is raised and the operator is not present (when enabled)
Close Sash will alternate on/off with the velocity reading. The Screen color will
change to Red.
Emergency -
will be displayed if the Emergency input is activated. Emergency will alternate on/off
with the velocity reading. The Screen color will change to Red.
Up/Down Arrow - will be displayed if Hi/Lo 2 Speed operation is enabled.
Off / Start up xx - will be displayed if Fan Stop operation is enabled. Off will be displayed
when the input is active and the Start Up timer will be displayed when the input is off.
2.4 Additional Alarms and Events
Mains Fail -
will be displayed if the power fails to the monitor (when enabled)
*Note – this is an optional extra feature that requires an additional battery unit
Low Temp -
will be displayed if the fume hood temperature drops below the low temp alarm
point (when enabled).
This display will alternate on/off with the velocity reading
*Note – this is an optional extra feature that requires an additional temperature sensor
High Temp -
will be displayed if the fume hood temperature rises above the high temp alarm
point (when enabled)
*Note – this is an optional extra feature that requires an additional temperature sensor
Содержание Air Alert 600
Страница 2: ...2 Issue 02 June 2017...
Страница 5: ...5 1 SAFETY...
Страница 10: ...10 2 Overview of the Airflow Monitor...
Страница 20: ...20 3 Functions Operation...
Страница 25: ...25 4 Components...
Страница 28: ...28 5 Installation...
Страница 33: ...33 6 Configuration Calibration...
Страница 37: ...37 7 Auxiliary Features Connections...
Страница 40: ...40 8 RS485 Coms Output...
Страница 47: ...47 9 Warranty...