![Kett HK-300-1 Скачать руководство пользователя страница 8](http://html1.mh-extra.com/html/kett/hk-300-1/hk-300-1_operating-manual_1958593008.webp)
8
2. Specifications
Measurement Principle
:
Electrical resistance
Applications
:
All paper
Calibration Curves *
1
:
See "Table 1"on the right
Measurement Range
:
See "Table 1" on the right
Measurement Precision
:
Standard error 0.4% (moisture values of 15% or lower)
Standard Method
:
JIS P8127 (copying paper, cardboard)
Display
:
Digital (LCD)
Resolution
:
0.1%
Operating Temperature Range
:
0 to 40˚C
Functions
:
User calibration curves (10 types),Data storage (250 data), Moisture value bias
adjustment (-9.9 to +9.9%), Average value display, Upper limit alarm setting, Auto
power off (automatically goes off in about 5 minutes)
Power Supply, Power Consumption
:
1.5 V batteries (AA alkaline) (6), approximately 0.45 W
Dimensions and Weight
:
110 mm (W) x 210 mm (D) x 50 mm (H), 0.5 kg
Included sensors
:
HK-300-1: grip sensor (cable included)
HK-300-2: rubber sensor (underlay, conducting rubber (2), wrench included)
HK-300-3: pressure sensor
Accessories
:
Shoulder strap, 1.5 V batteries (AA alkaline) (6), Carrying case, Operating manual,
Calibration curve list (Japanese and English) (1 each)
Options
:
Printer VZ-390, Printer cable VZC70