29
(3) Record the measurement results in count mode (calibration
curve number 10) as in Table 1. (Please copy and use the
"Calibration Curve Preparation Table" on p. 37.)
* When preparing the table, put the count values from the HK-300 in the
ascending order.
7-4. Inputting a Calibration Curve
(when measuring 5 sample points)
10 user calibration curves (calibration curve numbers 11 to 20)
can be input into this unit. Here, we describe an example of
entering and storing measurement results (content of "Table
1") in count mode (calibration curve number 10) into calibration
curve number "12" of the HK-300.
* Enter numerical input with the numeric keys. When the count value to
input is a 2-digit integer, or when the moisture value is a 1-digit, enter a
leading "0".
<Example> To input 9.9%
Press keys in the order of
* If you make a mistake in the numerical value to be input, press the
key, and then enter the correct numerical value.
<Table 1>
◆
Calibration curve number :
◆
Calibration curve name :
◆
Sample name
:
◆
Sensor name
:
◆
Measurement temperature :
HK-300 count value
(calibration curve number 10)
Moisture value per
a standard method
Sample
1
98
3.6%
Sample
2
319
5.7%
Sample
3
473
9.8%
Sample
4
665
16.9%
Sample
5
869
21.2%