* To input a calibration curve, the numerical values for points [1] through
[5] must be entered. For preparing a 2 to 4 point sample calibration
curve, repeatedly enter the upper limit values (count value = 999,
moisture value = 99.9). For example, when preparing a 2-point sample
calibration curve, enter the upper limit values (HK-300 = 999, standard
method = 99.9) for all of the measurement values for [Sample
3
] to
[Sample
5
].
* If newly entering a different calibration curve for a calibration curve
number (11-20) that has already been entered, the calibration curve
will be overwritten, and the previously entered calibration curve will
be modified. Because there may be times when a needed calibration
curve is erased by mistake, we recommend that calibration curve data
be saved once it has been prepared.