L.E.D Curing Light
Operator Manual
MADE IN USA
DESCRIPTION
The Demi
Plus
L.E.D. (Light Emitting Diode) curing light is a visible curing unit intended
for polymerization of light-cured materials by dental professionals.
IMPORTANT:
Batteries must be fully charged prior to use. Refer to the CHARGE THE
BATTERY(IES) section. There are no user serviceable parts inside. Do not open or
tamper with the batteries.
SPECIFICATIONS AND EQUIPMENT CLASS
AC Supply connection:
100-240V AC/0.8-0.4A 47-63 Hz
Power input:
12VA, nominal
Equipment class:
Class II
Safety:
IEC 60601-1
EMC (Electro-Magnetic Compliance):
IEC 60601-1-2
Protection from electric shock:
Type BF
Protection from ingress of liquids:
IXP0 (ordinary equipment)
Operation:
Continuous operation patient application,
duty cycle 20 seconds ON/1 minute OFF.
OPERATING ENVIRONMENT
Note:
Take caution when using this
product in the presence of a
fl
ammable anesthetic
mixture or an oxidizer such as oxygen or nitrous oxide.
·
Ambient temperature: 60˚F to 104˚F (16˚C to 40˚C)
·
Relative humidity: 10% to 80%, non-condensing
·
Atmospheric pressure: 0.5-atm to 1.0-atm (500 hPa to 1060 hPa)
TRANSPORT AND STORAGE ENVIRONMENT
·
Ambient temperature: -4˚F to 104˚F (-20˚C to 40˚C)
·
Relative humidity: 10% to 85%
·
Atmospheric pressure: 0.5 atm to 1.0 atm (500 hPa to 1060 hPa)
·
Battery performance: Handpiece will generate 500,
fi
ve-second cures with a fully
charged battery
·
Battery over-current protection: Electronic and Resettable polyfuse
·
Light Emitting Diodes: An array of four LEDs die-mounted on a substrate.
·
Output peak wavelength range: 450 - 470 nanometers
·
Baseline output: 1,100 mW/cm
2
·
Standard light guide(s): Part Number 902495, 13mm to 8mm Extended Light Guide
Unit handpiece dimensions:
Length: 9.25 in (23.5 cm); Width: 1.2 in (2.9 cm)
Charger dimensions:
Length: 6.0 in (16.0 cm); Width: 4.63 in (11.1 cm)
Unit weight:
Handpiece and battery: 5.5 oz (155 g); Battery charger: 12.4 oz (352 g)
SYMBOLS
Attention, consult accompanying documents:
DC power:
Waste Electrical and Electronic Equipment (WEEE):
SAFETY PRECAUTIONS
t
5IFöCFSPQUJDMJHIUHVJEFJTQSPWJEFEJOBOPOTUFSJMFDPOEJUJPOBOENVTUCF
sterilized prior to patient contact. For sterilization instructions, refer to the
MAINTENANCE: CLEANING/DISINFECTING/STERILIZATING section.
t
5IJTDVSJOHVOJUQSPEVDFTIJHIPVUQVUDVSJOHFOFSHZ"TJHOJöDBOUJODSFBTF
in curing energy is possible compared with equipment previously used. It is
important to observe the following precautions and procedures:
t
%POPUQMBDFMJHIUEJSFDUMZPOPSUPXBSETVOQSPUFDUFEHJOHJWBPSTLJO
t
"EKVTUDVSJOHUFDIOJRVFTJOBDDPSEBODFXJUIUIFJODSFBTFJODVSJOHFOFSHZ
t
%POPUMPPLEJSFDUMZBUUIFMJHIUFNJUUFEGSPNUIJTDVSJOHVOJU%POPUVTF
this device without suitable eye protection for the operator, assistant and
patient. Refer to the ACCESSORIES AND REPLACEMENT PARTS section for
suitable eye protection.
t
1FSTPOTIBWJOHBIJTUPSZPGQIPUPTFOTJUJWFSFBDUJPOTPSXIPBSFVTJOH
photosensitizing drugs should not be exposed to light from this unit.
t
/05&5BLFDBVUJPOXIFOVTJOHUIJT
product in the presence of a
fl
ammable
anesthetic mixture or an oxidizer such as oxygen or nitrous oxide.
t
"MM%FNFUSPONPEFMTIBWFCFFOEFTJHOFEUPNFFUXPSMEXJEFFMFDUSJDBMTBGFUZ
standards, including U.S., Canada, and Europe, to be safe and e
ff
ective for all
dental applications.
t
5IF%FNJ
Plus
requires special precautions regarding EMC and must be installed
and put into service according to the Guidance and Manufacturer’s Declaration
for EMC included in this manual.
t
1PSUBCMFBOENPCJMF3'DPNNVOJDBUJPOTFRVJQNFOUDBOBòFDUUIF%FNJ
Plus
.
Reference the Guidance and Manufacturer’s Declaration for EMC included in this
manual.
t
We are certain that you will receive many years of exceptional service from the
Demi
Plus
product. However, it is inevitable that at some time in the future you
will need to dispose of it. When that time comes, we encourage you to return it
to Kerr in its original packaging for environmentally sound recycling. Kerr will
be glad to bear the cost of the return shipping. Please contact your sales rep for
details. The Waste Electrical and Electronic Equipment (WEEE) Directive forbids
the disposal of waste electrical and electronic equipment as unsorted municipal
waste and requires that they be collected and recycled or disposed of separately.
Hazardous substances are present in electrical and electronic equipment and
present potential risks to human health and the environment if disposed of in
municipal land
fi
lls which are not designed to prevent migration of substances
into the soil and groundwater. By properly disposing of this waste, you will be
contributing to the advancement of reuse, recycling, and other forms of recovery,
as well as the prevention of pollution at a very local level.
NOTE: The dimensions in the
fi
gure represent the minimum patient environment (1.5 meters).
The Demi
Plus
charger base should not be stored within this patient environment.
ASSEMBLY/MOUNTING
The box contains:
Qty.
Description
1
Demi
Plus
curing light handpiece
1
Protective light shield
1
Plug-in power supply
1
Battery charger
1
Battery pack
1
Operator manual
1
Demi/Demi
Plus
disposable barrier