ILT-BA-e-0612
59
01 Reference weight determination of reference weight through weighing
Call up Percent weighing mode
Percent setup
M4
01 Reference
02 Decimal places
03 Checkweigh ing
04 Filling
05 Statistics
06 Run
0.0000
1 PLAC E
OFF
OFF
ON
g
Press
-button. Put on reference weight; if necessary,
tare first
Press
-key, the weight is taken as reference (100%).
Min. weight: readout d x 1000
Percent
100,0
%
0%
100%
L
REF =100.000
WGH=0.000
24.01.05
Now you can place the test objects onto the weighing
plate; the percentage is displayed
Percent
17,3
%
0%
100%
L
REF =100.000
WGH=0.000
24.01.05
02 Decimal places
Use the cursor (
) to select menu item
02 Decimal
places"
Press the
-key; the current menu item blinks.
Use the arrow keys, to enter the fractional digit, confirm
by pressing the
-key.
Press the
-key; the balance is now in percent
determination mode.
03 Check weighing
04 Filling
05 Statistics
06 Run
Use the cursor (
) to select menu item 03 - 06"
All other steps are the same as for combination Parts
counting /Statistics (chpt. 9.1.2)