ILT-BA-e-0612
21
Now move the adjusting weight to the centre of
the weighing plate, press
-key. If the
adjustment was successful, this will be followed
by automatic reset to the menu.
The
-key may be used to cancel the
adjustment.
Errors during adjustment or the use of an
incorrect adjusting weight will result in an error
message. Repeat adjustment.
Einstellungen
P 1 01
02
03
04
Ext.
test
Pr
calibr.
alibr.
User calibr.
C
int rep ort
********
********
********
1
f u n c t io n
f u n c t io n
o n
f u n c t io n
29.12.04
We recommend:
Carry out all adjustments as close as possible to the maximum load of the
balance. Info about test weights can be found on the Internet at: http://www.kern-
sohn.com
03 Cal. test
Here, deviation from the last adjustment is determined. This is only a check, i.e. no
values are changed.
Use the cursor (
) to select "
03 Cal. test
"
Einstellungen
P 1 01
02
03
04
Ext.
test
Pr
calibr.
alib r.
User calibr.
C
int report
********
********
********
1
fu n c tio n
fu n c tio n
o n
fu n c tio n
29.12.04
Press
-button
Calibration
Calibration check
Clear the pan
[Enter]
Press
-key, on the display appears the weight
value of the adjusting weight.
Calibration
Calibration check
weight measuring
Einstellungen
Calibration check
Load weight
[
]
12000.0g
Enter
Now set the adjusting weight in the centre of the
weighing plate. Press
-key; calibration test will
be executed automatically
Kalibrierung
Kalibrierungstest
Bitte warten
The result is displayed
Calibration
Check results
Cal.: 180.1354
Act.: 180.1438
Diff.: - 0.0084