ILT-BA-e-0612
55
9.1.3 Checkweighing (weighing with tolerance range)
Overview of display:
1. Current weighing value
2. Weighing assistance
3. Operating mode
4. Nominal value
5. Upper limit (Hi), lower limit (Lo)
6. Tolerance marker
too light
nominal
value
too heavy
Operation
Display
Call up menu item "
M2 Checkweighing
"
(see chap. 9.1)
Press
-button
01 - 02 Lower/upper limit entering limit values
Use the cursor (
) to select
01 Lower limit or
02 Upper limit
Press the
-key; the current menu item blinks.
Use the arrow keys, to enter the limits and confirm by
pressing the
-key
Notice: First enter upper limit.