ILT-BA-e-0612
24
Unlock the balance for the adjusting process:
Operation
Display
Remove the seal of approval
(Attention: Verification no longer valid!)
Switch off the balance
Switch on the balance whilst keeping the
adjusting switch pressed down
As soon as the pictograph for balance appears,
release the adjusting switch.
Wait until the balance is in weighing mode
Press
-button
13 :47:32
0.0000
Weighing
29.12.04
P1 Calibration
P2 GLP
P3 Date/Time
P4 Readout
P5 RS-232
P6 Printouts
P7 Units
P8 Modes
P9 Globals
Setup
2.02.05
Access to service menu
P0 factory settings
:
Press
-button
P0 Factory
n
P2 GLP
P3 Date/Time
P4 Readout
P5 RS-232
P6 Printouts
P7 Units
P8 Modes
P9 Globals
P1 Calibratio
Setup
2.02.05
Press
-key, selection of parameters appears
Trigger parameter
18 LFT
by pressing the
-key and select
0
setting. Your balance is
now unlocked for the adjusting process and all
weighing units are activated.
Einstellungen
P
0
14 Autozero delay 6 2.0 s
15 Stable range 10 1.0 d
16 Stable speed 4 0.8 s
17 Filter range 7 100 d
18 LFT 0 off
Confirm and store by pressing the
-key
(see chpt. 8.1)
To reset the balance to the calibrated setting,
select setting "
1
for parameter
18 LFT
. Now only
the metric weighing units (g, kg) are activated.