4. EVERYDAY USE OF THE VALVE
This chapter gives an introduction to how the sampling valve works in different operating conditions,
such as the cleaning of the entire production line before starting a new batch (chapter 4.1) and the
cleaning of the valve between each sample during the batch production (chapters 4.2-4.4)
For specific operator instructions please refer to the chapter 5. “VALVE OPERATIONS”.
4.1 Batch change cleaning
Before every new production batch the sampling valve is cleaned and disinfected/sterilised together
with the tank or vessel or the entire production line.
Make sure the valve is in its OPEN position during the initial line CIP to allow cleaning of the valve seat
and the membrane contact surface.
Also allow CIP fluid, disinfectant or steam to flow through the inlet and outlet hose pieces. If the valve
is fitted to a tank, which is spray cleaned there will be insufficient flow of CIP liquid through the valve.
Therefore connect CIP hoses to the two ports for CIP fluids to flow through the valve into the tank (see
illustration below).
If the valve is fitted to a pipe there will usually be enough pressure and flow during the CIP process for
the CIP fluid to flow through the valve.
PIPE
TANK
Remember to close the valve after the final rinse and prior to starting up the next production batch.
4.2 Chemical cleaning, CIP
During production and prior to sampling, cleaning takes place with the valve closed and involves the
following stages:
1. Pre-rinse
Flushing with water to mechanically remove product residues
2. Clean
Applying a detergent to remove remaining visible product residues
3. Final rinse
Rinse with clean water to remove all traces of detergents
Usually this procedure is followed by disinfection (see below), but for some application CIP might be
sufficient. It depends on your (microbiological) requirements, the detergents applied and the process
media to clean for. Consult your supplier of CIP fluids.
In some cases where the process media is for instance water, CIP might not even be necessary and you
may go directly to disinfection.
4.3 Chemical Disinfection
Disinfection takes place with the valve closed and involves the following stages of which the first 3 are
identical to CIP:
1. Pre-rinse
Flushing with water to mechanically remove product residues
2. Clean
Applying a detergent to remove remaining visible product residues
Содержание W9
Страница 1: ...DON T GAMBLE WITH YOUR SAMPLE W9 SAMPLING VALVE User Manual ...
Страница 4: ...KEOFITT W9 USER MANUAL V 4 PAGE 4 ...
Страница 36: ...KEOFITT W9 USER MANUAL V 4 PAGE 36 12 5 Keofitt valve type Varivent 12 6 Keofitt valve type thread ...
Страница 44: ...KEOFITT W9 USER MANUAL V 4 PAGE 44 16 MEMBRANES 16 1 Silicone membrane art no 600051 ...
Страница 45: ...KEOFITT W9 USER MANUAL V 4 PAGE 45 16 2 Silicone membrane for Micro Port art no 600251 ...
Страница 46: ...KEOFITT W9 USER MANUAL V 4 PAGE 46 16 3 EPDM membrane art no 600052 ...
Страница 47: ...KEOFITT W9 USER MANUAL V 4 PAGE 47 16 4 EPDM membrane for Micro Port art no 600252 ...
Страница 48: ...KEOFITT W9 USER MANUAL V 4 PAGE 48 16 5 FFKM membrane art no 600053 ...
Страница 49: ...KEOFITT W9 USER MANUAL V 4 PAGE 49 16 6 PTFE membrane art no 850055 ...