K-O2 Oxygen Sensor 

5/20/2022 

Kele, Inc. • 3300 Brother Blvd. • Memphis, TN  38133 

Page 10 

 

www.kele.com

 

 

5.3  Z

ERO 

C

ALIBRATION 

P

ROCEDURE

 

For maximum accuracy below 18%, the Zero calibration procedure must be done before the Span calibration. 

The progress and status of the calibration process is indicated by the color and flash-state of the front cover status 
LED (see Table 10).   

Apply the nitrogen (zero) calibration gas to the sensor using the included calibration cap.  Ensure that gas is flowing 
to the sensor, then press and hold the ’ZERO’ button (see Figure 8) for 3 seconds until the front cover LED starts 
blinking YELLOW, indicating that gas sampling is in progress.   

1.

 

Ensure that the calibration adapter remains correctly seated and calibration gas continues to flow for the 
2 minute sampling period.   

2.

 

At the end of the sampling period, the sensor’s status LED blinks GREEN if the sampling was successful or 
RED if not.  

3A.  If successful (blinking GREEN): 

The gas sampling completed successfully. Turn off the calibration gas flow, remove the calibration cap 
then 

press and hold

 the ‘ZERO’ calibration button until the LED blinks GREEN/YELLOW indicating that the 

calibration gas has been removed, the calibration has been applied and the unit is in standby for two 
minutes while the sensor re-equilibrates to the ambient atmosphere before normal operation resumes.  
The calibration is complete when the status LED returns to steady GREEN. 

OR 

       3B.  If NOT successful (blinking RED):  

The most likely cause of zero calibration sampling failure is insufficient gas flow or leaks around the 
calibration adapter failing to completely immerse the sensor in nitrogen.  Verify that calibration gas is still 
flowing at the required rate (pressure gauge reads 10 psi) and the calibration adapter is properly 
positioned.   

The calibration sampling can be re-started

 while the LED is blinking RED by again 

pressing and holding

 

the ‘ZERO’ button until the LED blinks YELLOW, then return to step 1 above. 

To exit the zero-calibration routine preserving the original calibration: 

 turn off the calibration gas flow 

and remove the calibration adapter, then 

press and quickly release

 the ‘ZERO’ button.  The status LED will 

blink RED/YELLOW indicating that the calibration gas has been removed, the original calibration has been 
kept and the unit is in standby for two minutes while the sensor re-equilibrates to the ambient 
atmosphere before normal operation resumes.  The original calibration is completely restored when the 
status LED returns to steady GREEN.  

 

 

 

Содержание K-O2

Страница 1: ...mphis TN 38133 Revision 5 20 2022 www kele com 90 0007 00A K O2 xx Oxygen Sensor Transmitter and Two Stage Alarm Controller User s Manual K O2 H5 5 year sensor K O2 H10 10 year sensor K O2 S5 5 year s...

Страница 2: ...ional Description 5 4 1 Special Modes 6 4 2 Warning Ventilation and Alarm Conditions 7 4 3 Setting Ventilation and Alarm Thresholds 7 4 4 Concentration Reporting 7 5 Sensor Calibration 7 5 1 Calibrati...

Страница 3: ...sor life K O2 H5 Lockable hinged enclosure with 10 year sensor life K O2 H10 Table 1 K O2 family Part numbers All K O2 xx models are shipped with either 5 year life K O2 x5 or 10 year life K O2 x10 fa...

Страница 4: ...ply separate transformer not required Power Consumption 5W Control Relays 2 separate SPDT line voltage capable relays for warning ventilation and alarm outputs UL rated 10 Amps max at 120 277 VAC or 3...

Страница 5: ...water entry DO NOT USE THE VENT HOLES FOR WIRE ENTRY 1 This unit is designed to mount to a rigid vibration free surface near the middle of the area to be monitored about 5 feet above the floor 2 It s...

Страница 6: ...of the terminal and returns to the terminal The oxygen sensor output is provided at the terminal highlighted in Figure 3 Analog output connection has polarity as labeled on the controller silkscreen...

Страница 7: ...to the COM location of screw terminal 5 Plug the relay output screw terminal back into the correct location on controller board 3 3 POWER CONNECTION The K O2 has a fully isolated unpolarized power inp...

Страница 8: ...a second contact closure is operated typically to trigger an alarm The gas sensor is a calibrated module that can be replaced with minimal effort when it reaches end of life EOL while leaving the main...

Страница 9: ...ormation When the concentration of oxygen falls below its configured warning ventilation threshold the WARNING VENTILATION relay output is activated When the concentration falls below the alarm thresh...

Страница 10: ...adequate and requires no calibration gas All that s required is the certainty of fresh air around the sensor For accuracy at lower oxygen percentages a zero calibration before the span calibration is...

Страница 11: ...rictor fitted to the narrow end of the cal cap at a pressure of 10 psi 5 2 CALIBRATION GAS CONNECTION A schematic of the calibration gas tubing connection between the regulator and the calibration cap...

Страница 12: ...tion gas has been removed the calibration has been applied and the unit is in standby for two minutes while the sensor re equilibrates to the ambient atmosphere before normal operation resumes The cal...

Страница 13: ...turns to steady GREEN OR 3B If NOT successful blinking RED The most likely causes of span gas sampling failure are Insufficient gas flow or leaks around the calibration adapter not completely immersin...

Страница 14: ...n in To replace a sensor module with a new factory calibrated one follow the steps below 1 Open the controller s front cover 2 Unplug the controllers power connector refer to Figure 6 3 Unplug the sen...

Страница 15: ...patible with those Products Unless KELE agrees otherwise to obtain service under this warranty Buyer must pack any nonconforming Product carefully and ship it postpaid or freight prepaid to Kele Inc a...

Страница 16: ...tect common failures If a failure is detected the front cover status LED will blink slowly RED the warning relay will be activated and the concentration reporting analog output will stay at 4 mA until...

Отзывы: