Kele K-O2 Скачать руководство пользователя страница 10

K-O2 Oxygen Sensor 

5/20/2022 

Kele, Inc. • 3300 Brother Blvd. • Memphis, TN  38133 

Page 8 

 

www.kele.com

 

 

S

ENSOR 

C

ALIBRATION

 

The sensitivity of the galvanic oxygen sensor used on the K-O2 series decreases as the sensor ages.  Over the 
sensor’s lifetime its accuracy decreases by about 30%.  Without intervening calibrations, the sensor will typically 
indicate about 14.7 % oxygen concentration in fresh air after 5 (for the K-O2-x5) or 10 (for the K-O2-x10) years.   

Required calibration frequency depends on the accuracy requirement of the application.  To maintain the accuracy 
specified in Table 5 over the full operating range of the K-O2 series, a full calibration interval of 6 months is 
recommended.  Annual calibration will typically maintain an accuracy within about 0.5% O

2

 (for K-O2-x5) and 

about 0.3% O

2

 (for K-O2-x10).   

For best accuracy, the full two-step calibration process that provides the sensor module with oxygen-free ‘zero’ 
gas, and then a 21% ‘span’ gas is required.  Two calibration buttons (

ZERO

 and 

SPAN

) are provided on the main 

board to initiate each calibration operation as shown in Figure 8. 

For applications between 18% and 21% oxygen, a span-only calibration is often adequate and requires no 
calibration gas.  All that’s required is the certainty of fresh air around the sensor.  For accuracy at lower oxygen 
percentages, a zero calibration before the span calibration is recommended. 

To perform a gas-less span calibration

: follow the procedure in section 5.4, ignoring all instructions regarding the 

application or removal of calibration gas or fittings. 

The ‘sensor expired’ test will be performed at the end of a span calibration.  

If the sensor’s sensitivity has fallen 

below manufacturer’s end-of-life specification, the K-O2 goes into Sensor Expired mode with the front cover LED 
slowly blinking RED, the analog output at a constant 4 mA and the warning/ventilation relay activated.  The 
oxygen sensor is no longer operational and must be replaced (See section 6). 

The status of the calibration process is indicated by the flash pattern of the front cover LED as shown below.  

Blinking Green 

Successful sampling.  Waiting for user to confirm cal 

gas removal. 

Blinking Red 

Failed calibration attempt.  Waiting for user to 

acknowledge with either a re-try or an exit. 

Green/Yellow 

During ambient equilibration period after successful 

calibration.  

New calibration is applied.

 

Red/Yellow 

 

During ambient equilibration period after failed 

sampling.  

Old calibration is unchanged.

 

Table 10:  

Meaning of Status LED Blink Patterns During Calibration.

 

 

 

Содержание K-O2

Страница 1: ...mphis TN 38133 Revision 5 20 2022 www kele com 90 0007 00A K O2 xx Oxygen Sensor Transmitter and Two Stage Alarm Controller User s Manual K O2 H5 5 year sensor K O2 H10 10 year sensor K O2 S5 5 year s...

Страница 2: ...ional Description 5 4 1 Special Modes 6 4 2 Warning Ventilation and Alarm Conditions 7 4 3 Setting Ventilation and Alarm Thresholds 7 4 4 Concentration Reporting 7 5 Sensor Calibration 7 5 1 Calibrati...

Страница 3: ...sor life K O2 H5 Lockable hinged enclosure with 10 year sensor life K O2 H10 Table 1 K O2 family Part numbers All K O2 xx models are shipped with either 5 year life K O2 x5 or 10 year life K O2 x10 fa...

Страница 4: ...ply separate transformer not required Power Consumption 5W Control Relays 2 separate SPDT line voltage capable relays for warning ventilation and alarm outputs UL rated 10 Amps max at 120 277 VAC or 3...

Страница 5: ...water entry DO NOT USE THE VENT HOLES FOR WIRE ENTRY 1 This unit is designed to mount to a rigid vibration free surface near the middle of the area to be monitored about 5 feet above the floor 2 It s...

Страница 6: ...of the terminal and returns to the terminal The oxygen sensor output is provided at the terminal highlighted in Figure 3 Analog output connection has polarity as labeled on the controller silkscreen...

Страница 7: ...to the COM location of screw terminal 5 Plug the relay output screw terminal back into the correct location on controller board 3 3 POWER CONNECTION The K O2 has a fully isolated unpolarized power inp...

Страница 8: ...a second contact closure is operated typically to trigger an alarm The gas sensor is a calibrated module that can be replaced with minimal effort when it reaches end of life EOL while leaving the main...

Страница 9: ...ormation When the concentration of oxygen falls below its configured warning ventilation threshold the WARNING VENTILATION relay output is activated When the concentration falls below the alarm thresh...

Страница 10: ...adequate and requires no calibration gas All that s required is the certainty of fresh air around the sensor For accuracy at lower oxygen percentages a zero calibration before the span calibration is...

Страница 11: ...rictor fitted to the narrow end of the cal cap at a pressure of 10 psi 5 2 CALIBRATION GAS CONNECTION A schematic of the calibration gas tubing connection between the regulator and the calibration cap...

Страница 12: ...tion gas has been removed the calibration has been applied and the unit is in standby for two minutes while the sensor re equilibrates to the ambient atmosphere before normal operation resumes The cal...

Страница 13: ...turns to steady GREEN OR 3B If NOT successful blinking RED The most likely causes of span gas sampling failure are Insufficient gas flow or leaks around the calibration adapter not completely immersin...

Страница 14: ...n in To replace a sensor module with a new factory calibrated one follow the steps below 1 Open the controller s front cover 2 Unplug the controllers power connector refer to Figure 6 3 Unplug the sen...

Страница 15: ...patible with those Products Unless KELE agrees otherwise to obtain service under this warranty Buyer must pack any nonconforming Product carefully and ship it postpaid or freight prepaid to Kele Inc a...

Страница 16: ...tect common failures If a failure is detected the front cover status LED will blink slowly RED the warning relay will be activated and the concentration reporting analog output will stay at 4 mA until...

Отзывы: