Model 6485 Picoammeter Instruction Manual
Measurement Concepts
2-15
SCPI programming — zero check and zero correct
A) SYSTem:ZCORrect:ACQuire
The zero correct value can only be acquired while zero check is enabled and zero correct
state is off. The internal offset measured at that moment will become the correction value.
Zero correction can then be applied and zero check disabled. This acquire method makes it
convenient if you need to re-zero the instrument often.
The following command sequence uses the acquire method to zero correct the 200
µ
A
range:
*RST
' Set instrument to known default
conditions in one-shot trigger mode.
SYST:ZCH ON
' Enable zero check.
CURR:RANG 2E-4
' Set instrument to 200uA range.
INIT
' Trigger one reading.
SYST:ZCOR:ACQ
' Acquire zero correct value.
SYST:ZCH OFF
' Disable zero check.
SYST:ZCOR ON
' Perform zero correction.
The INITiate command in the above sequence is used to trigger a reading. This reading is
the offset that is acquired as the zero correct value. See
INITiate.
NOTE
Sending the :ACQuire command while zero check is disabled will result in an
error. The command will not be executed.
Table 2-4
SCPI commands — zero check and zero correct
Commands
Description
Default
Ref
DDC
SYSTem
:ZCHeck
[:STATe] <b>
:ZCORrect
[:STATe] <b>
:ACQuire
INITiate
SYSTem Subsystem:
Zero check:
Enable or disable zero check.
Zero correct:
Enable or disable zero correct.
Acquire a new zero correct value.
Trigger a reading.
ON
OFF
OFF
A
A
B
B
C0 (Off)
C1 (On)
Содержание 6485
Страница 1: ...Model 6485Picoammeter Instruction Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E ...
Страница 20: ......
Страница 24: ......
Страница 44: ...1 20 Getting Started Model 6485 Picoammeter Instruction Manual ...
Страница 66: ...3 6 Measurements Model 6485 Picoammeter Instruction Manual ...
Страница 138: ...9 20 Remote Operation Model 6485 Picoammeter Instruction Manual ...
Страница 159: ...11 CommonCommands ...
Страница 164: ...11 6 Common Commands Model 6485 Picoammeter Instruction Manual ...
Страница 165: ...12 SCPISignalOriented MeasurementCommands ...
Страница 190: ...14 12 SCPI Reference Tables Model 6485 Picoammeter Instruction Manual ...
Страница 214: ...16 14 Calibration Model 6485 Instruction Manual ...
Страница 219: ...A Specifications ...
Страница 221: ...B StatusandErrorMessages ...
Страница 227: ...C GeneralMeasurement Considerations ...
Страница 233: ...D DDCEmulationCommands ...
Страница 247: ...E ExamplePrograms ...
Страница 250: ...E 4 Example Programs Model 6485 Picoammeter Instruction Manual ...
Страница 251: ...F IEEE 488BusOverview ...
Страница 266: ...F 16 IEEE 488 Bus Overview Model 6485 Picoammeter Instruction Manual ...
Страница 267: ...G IEEE 488andSCPI ConformanceInformation ...
Страница 311: ......