11-4
Common Commands
Model 6485 Picoammeter Instruction Manual
C) SAV <NRf> — save
Save present setup in memory
RCL <NRf> — recall
Return to setup stored in memory
Parameters
0 = Memory location 0
1 = Memory location 1
2 = Memory location 2
Use the *SAV command to save the present instrument setup configuration in memory for
later recall. Any control affected by *RST can be saved by the *SAV command. The *RCL
command is used to restore the instrument to the saved setup configuration. Three setup
configurations can be saved and recalled.
Model 6485 ships from the factory with SYSTem:PRESet defaults loaded into the avail-
able setup memory. If a recall error occurs, the setup memory defaults to the SYS-
Tem:PRESet values.
Programming example:
*SAV 2
' Save present setup in memory location 2.
*RST
' Set 6485 to RST defaults.
*RCL 2
' Return (recall) 6485 to setup stored in
memory
location 2.
D) RST — reset
Return Model 6485 to RST defaults
When the *RST command is sent, Model 6485 performs the following operations:
1.
Returns Model 6485 to the RST default conditions (see “Default” column of SCPI
tables).
2.
Cancels all pending commands.
3.
Cancels response to any previously received *OPC and *OPC? commands.
E) TRG — trigger
Send bus trigger to Model 6485
Use the *TRG command to issue a GPIB trigger to Model 6485. It has the same effect as a
group execute trigger (GET).
Use the *TRG command as an event to control operation. Model 6485 reacts to this trigger
if BUS is the programmed arm control source. The control source is programmed from the
TRIGger subsystem.
NOTE
Details on triggering are covered in
Содержание 6485
Страница 1: ...Model 6485Picoammeter Instruction Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E ...
Страница 20: ......
Страница 24: ......
Страница 44: ...1 20 Getting Started Model 6485 Picoammeter Instruction Manual ...
Страница 66: ...3 6 Measurements Model 6485 Picoammeter Instruction Manual ...
Страница 138: ...9 20 Remote Operation Model 6485 Picoammeter Instruction Manual ...
Страница 159: ...11 CommonCommands ...
Страница 164: ...11 6 Common Commands Model 6485 Picoammeter Instruction Manual ...
Страница 165: ...12 SCPISignalOriented MeasurementCommands ...
Страница 190: ...14 12 SCPI Reference Tables Model 6485 Picoammeter Instruction Manual ...
Страница 214: ...16 14 Calibration Model 6485 Instruction Manual ...
Страница 219: ...A Specifications ...
Страница 221: ...B StatusandErrorMessages ...
Страница 227: ...C GeneralMeasurement Considerations ...
Страница 233: ...D DDCEmulationCommands ...
Страница 247: ...E ExamplePrograms ...
Страница 250: ...E 4 Example Programs Model 6485 Picoammeter Instruction Manual ...
Страница 251: ...F IEEE 488BusOverview ...
Страница 266: ...F 16 IEEE 488 Bus Overview Model 6485 Picoammeter Instruction Manual ...
Страница 267: ...G IEEE 488andSCPI ConformanceInformation ...
Страница 311: ......