Model 6485 Picoammeter Instruction Manual
DISPlay, FORMat, and SYSTem
13-9
A) SYSTem:PRESet
Returns the instrument to states optimized for front panel operation. SYSTem:PRESet
defaults are listed in the SCPI tables in
.
B) SYSTem:TIME:RESet
Resets the absolute timestamp to 0 seconds. The timestamp also resets when power is
cycled, or after the instrument is on for 99,999.99 seconds. The TRACe:TSTamp:FORMat
command is used to select the absolute timestamp. See
(Buffer) for details.
C) SYSTem:POSetup <name>
Parameters
RST
Power-up to *RST defaults
PRESet
Power-up to SYSTem:PRESet defaults
SAVx
Power-up to setup stored in memory (x = memory location
0, 1 or 2)
The *RST and SYSTem:DEFaults are listed in the SCPI tables in the
. A setup
is saved in memory using the *SAV command. See
details.
D) SYSTem:VERSion
Read the version of the SCPI standard being used by Model 6485. Example response mes-
sage: 1996.0.
E) SYSTem:KEY <NRf>
Parameters
1 = CONFIG/LOCAL key
17 = MENU key
2 = MEDN key
18 = COMM key
3 = AVG key
19 = DISP key
4 = MX + B key
20 = TRIG key
5 = M/X + B key
21 = HALT key
6 = LOG key
22 = DIGITS key
7 = REL key
23 = RATE key
8 = ZCHK key
24 = Cursor left arrow key
9 = -------
25 = -------
10 = -------
26 = SAVE key
11 = RANGE up arrow key
27 = SETUP key
12 = AUTO key
28 = STORE key
13 = RANGE down arrow key
29 = RECALL key
14 = ENTER key
30 = LIMIT key
15 = Cursor right arrow key
31 = AZERO key
16 = ZCOR key
32 = EXIT key
This command is used to simulate front panel key presses. For example, send the follow-
ing command to simulate pressing the “MEDN” key: SYSTem:KEY 2. The key-press
codes are also shown in
.
Содержание 6485
Страница 1: ...Model 6485Picoammeter Instruction Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E ...
Страница 20: ......
Страница 24: ......
Страница 44: ...1 20 Getting Started Model 6485 Picoammeter Instruction Manual ...
Страница 66: ...3 6 Measurements Model 6485 Picoammeter Instruction Manual ...
Страница 138: ...9 20 Remote Operation Model 6485 Picoammeter Instruction Manual ...
Страница 159: ...11 CommonCommands ...
Страница 164: ...11 6 Common Commands Model 6485 Picoammeter Instruction Manual ...
Страница 165: ...12 SCPISignalOriented MeasurementCommands ...
Страница 190: ...14 12 SCPI Reference Tables Model 6485 Picoammeter Instruction Manual ...
Страница 214: ...16 14 Calibration Model 6485 Instruction Manual ...
Страница 219: ...A Specifications ...
Страница 221: ...B StatusandErrorMessages ...
Страница 227: ...C GeneralMeasurement Considerations ...
Страница 233: ...D DDCEmulationCommands ...
Страница 247: ...E ExamplePrograms ...
Страница 250: ...E 4 Example Programs Model 6485 Picoammeter Instruction Manual ...
Страница 251: ...F IEEE 488BusOverview ...
Страница 266: ...F 16 IEEE 488 Bus Overview Model 6485 Picoammeter Instruction Manual ...
Страница 267: ...G IEEE 488andSCPI ConformanceInformation ...
Страница 311: ......