Model 6485 Picoammeter Instruction Manual
DDC Emulation Commands
D-13
The various bits in the status byte are described below:
Reading Overflow
— Set when an overrange input is applied to the instrument.
Data Store Full
— Set when the number of readings stored is equal to the buffer size.
Reading Done
— Set when the instrument has completed the present conversion and is
ready to take another reading.
Busy
— The instrument is still executing a prior command and is not yet ready to accept a
new command.
IDDCO
— An illegal command option such as R8 has been sent. This bit is cleared when
the status byte is read.
IDDC
— An illegal command will set this bit. For example, S1 is illegal since no such let-
ter exists in the command set. The IDDC bit will be cleared on a reading of the status byte.
Not in remote
— Model 6485 in local mode of operation.
NOTE
Once the Model 6485 has generated an SRQ, its status byte must be read to
clear the SRQ line. Otherwise, the instrument will continuously assert SRQ.
The Model 6485 may be programmed to generate an SRQ for more than one
condition simultaneously.
Table D-2
Status byte and mask interpretation
Bit
Bit 5=0 (Data Conditions)
Bit 5=1 (Error Conditions)
0 (LSB)
Overflow
IDDCO
1
Data Store Full
IDDC
2
N/A
No remote
3
Reading done
N/A
4
Busy
N/A
5
Data
Error
6
SRQ
SRQ
7
N/A
N/A
Содержание 6485
Страница 1: ...Model 6485Picoammeter Instruction Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E ...
Страница 20: ......
Страница 24: ......
Страница 44: ...1 20 Getting Started Model 6485 Picoammeter Instruction Manual ...
Страница 66: ...3 6 Measurements Model 6485 Picoammeter Instruction Manual ...
Страница 138: ...9 20 Remote Operation Model 6485 Picoammeter Instruction Manual ...
Страница 159: ...11 CommonCommands ...
Страница 164: ...11 6 Common Commands Model 6485 Picoammeter Instruction Manual ...
Страница 165: ...12 SCPISignalOriented MeasurementCommands ...
Страница 190: ...14 12 SCPI Reference Tables Model 6485 Picoammeter Instruction Manual ...
Страница 214: ...16 14 Calibration Model 6485 Instruction Manual ...
Страница 219: ...A Specifications ...
Страница 221: ...B StatusandErrorMessages ...
Страница 227: ...C GeneralMeasurement Considerations ...
Страница 233: ...D DDCEmulationCommands ...
Страница 247: ...E ExamplePrograms ...
Страница 250: ...E 4 Example Programs Model 6485 Picoammeter Instruction Manual ...
Страница 251: ...F IEEE 488BusOverview ...
Страница 266: ...F 16 IEEE 488 Bus Overview Model 6485 Picoammeter Instruction Manual ...
Страница 267: ...G IEEE 488andSCPI ConformanceInformation ...
Страница 311: ......