Section 1: Verification procedures 3390 50 kHz Arbitrary Waveform / Function Generator Verification and Adjustment Manual
1-18
3390-905-01Rev. B / May 2016
ManualAdjustment.exe: Software operation
1.
Click
Begin
to execute the manual calibration. If the AWG manual calibration function is locked,
enter the correct secure code.
Most power meters require an attenuator or special power head to measure the +20 dB output. The
default security code uses "Model Name" (the first letter must be an alphabetic character). If the first
letter of "Model Name" is not alphabetic, the system will pick the first and last letter of the company
name for the header, plus "Model Name." For example, the Keithley 3390 would be "KY3390".
Figure 8: ManualAdjustment.exe secure code box
2.
Follow the instructions on the Manual Adjustment dialog box to correctly connect the cable.
Figure 9: ManualAdjustment.exe connect AWG output table