Section 1: Verification procedures 3390 50 kHz Arbitrary Waveform / Function Generator Verification and Adjustment Manual
1-8
3390-905-01Rev. B / May 2016
Checking 0 dB range flatness
This test checks the high frequency AC amplitude flatness above 100 KHz on the 0 dB attenuator
range. Flatness is relative to 1 KHz.
Figure 3: Power meter connections