3390 50 kHz Arbitrary Waveform / Function Generator Verification and Adjustment Manual Section 1: Verification procedures
3390-905-01Rev. B / May 2016
1-11
Ch20 db range flatness
This procedure checks the high frequency AC amplitude flatness above 100 kHz on the +20 dB
attenuator range. Flatness is relative to 1 kHz.
1. Connect the power meter to measure the output voltage of the instrument as shown below.
Figure 4: Power meter connections