Section 1: Verification procedures 3390 50 kHz Arbitrary Waveform / Function Generator Verification and Adjustment Manual
1-6
3390-905-01Rev. B / May 2016
AC amplitude verification for high impedance
This procedure checks the AC amplitude output accuracy at a frequency of 1 kHz, and establishes
reference measurements for the higher frequency flatness verification procedures.
1. Set the DMM to measure V
RMS
volts. Connect the DMM as shown below.
Figure 2: Power meter conenctions