Model 2750 Multimeter/Switch System Service Manual
Performance Verification
1-23
770
0 V
eri
fication
2.
Install the Model 7700 in Slot 1 of the Model 2750, then turn on the power, and
allow the unit to warm up for two hours before proceeding. Be sure the front panel
INPUTS switch is set to the REAR position.
3.
Select the AC volts function by pressing the ACV key. Close Channel 1 by press-
ing the CLOSE key, then ENTER for “SINGLE” channel, and then keying in 101.
4.
Set the Model 2750 for the 100mV range; make sure that REL is disabled.
5.
Source 1kHz and 50kHz AC voltages for each of the ranges summarized in
, and make sure that the respective Model 2750 readings fall within
stated limits.
6.
Press the OPEN key to open Channel 1.
Table 1-11
Plug-in module ACV reading limits
ACV
Range
Applied AC
Voltage
1kHz Reading Limits
(1 year, 18°C to 28°C)
50kHz Reading Limits
(1 year, 18°C to 28°C)
100mV
1V
10V
100V
750V
100.0000mV
1.000000V
10.00000V
100.0000V
300.000V*
99.910 to 100.090mV
0.99910 to 1.00090V
9.9910 to 10.0090V
99.910 to 100.090V
299.60 to 300.40V
99.830 to 100.170mV
0.99830 to 1.00170V
9.98300 to 10.0170V
99.830 to 100.170V
299.27 to 300.73V
*If the 5725A amplifier is not available, change the 300V @ 50kHz step to 220V @ 50kHz. Reading limits
for 220V @ 50kHz = 219.36 to 220.64V.
Содержание 2750
Страница 13: ...1 PerformanceVerification...
Страница 48: ...1 36 Performance Verification Model 2750 Multimeter Switch System Service Manual...
Страница 49: ...2 Calibration...
Страница 74: ...2 26 Calibration Model 2750 Multimeter Switch System Service Manual...
Страница 75: ...Routine Maintenance 3 RoutineMaintenance...
Страница 86: ...3 12 Routine Maintenance Model 2750 Multimeter Switch System Service Manual...
Страница 87: ...Troubleshooting 4 Troubleshooting...
Страница 111: ...Disassembly 5 Disassembly...
Страница 118: ...5 8 Disassembly Model 2750 Multimeter Switch System Service Manual...
Страница 119: ...Replaceable Parts 6 ReplaceableParts...
Страница 135: ...Specifications A AccuracyandOptimization...
Страница 140: ...A 6 Accuracy and Optimization Model 2750 Multimeter Switch System Service Manual...
Страница 141: ...Calibration Reference B CalibrationReference...
Страница 167: ...Calibration Program C CalibrationProgram...
Страница 178: ......
Страница 179: ......