Model 2750 Multimeter/Switch System Service Manual
Calibration
2-23
Mod
el 77
00 Cal
ibr
ation
6.
Perform the entire remote comprehensive calibration procedure discussed earlier in
this section. (See “
” earlier in this section.)
7.
Connect the function generator to the Model 2750 INPUT jacks as shown in
. Select the front input jacks with the INPUTS switch.
8.
Set the generator to output a 1V RMS, 3Hz sine wave, then send the following
command:
:CAL:PROT:AC:STEP14 <Cal_voltage>
Here <Cal_voltage> is the actual 3Hz generator signal amplitude you measured
previously.
9.
Set the generator to output a 1V RMS, 1kHz sine wave, then send the following
command:
:CAL:PROT:AC:STEP15 1E3
10.
Send the following commands to set calibration dates, save calibration, and lock
out calibration:
:CAL:PROT:DATE <year>, <month>, <day>
:CAL:PROT:NDUE <year>, <month>, <day>
:CAL:PROT:SAVE
:CAL:PROT:LOCK
Model 7700 calibration
The following procedures calibrate the temperature sensors on the Model 7700 plug-in
modules.
NOTE
For additional information about the Keithley modules, refer to the module
documentation.
Recommended test equipment
In order to calibrate the Model 7700, you will need equipment summarized in
Table 2-14
Recommended equipment for Model 7700 calibration
Digital Thermometer:
18 to 28°C, ±0.1°C
Keithley 7797 Calibration/Extender Board
Содержание 2750
Страница 13: ...1 PerformanceVerification...
Страница 48: ...1 36 Performance Verification Model 2750 Multimeter Switch System Service Manual...
Страница 49: ...2 Calibration...
Страница 74: ...2 26 Calibration Model 2750 Multimeter Switch System Service Manual...
Страница 75: ...Routine Maintenance 3 RoutineMaintenance...
Страница 86: ...3 12 Routine Maintenance Model 2750 Multimeter Switch System Service Manual...
Страница 87: ...Troubleshooting 4 Troubleshooting...
Страница 111: ...Disassembly 5 Disassembly...
Страница 118: ...5 8 Disassembly Model 2750 Multimeter Switch System Service Manual...
Страница 119: ...Replaceable Parts 6 ReplaceableParts...
Страница 135: ...Specifications A AccuracyandOptimization...
Страница 140: ...A 6 Accuracy and Optimization Model 2750 Multimeter Switch System Service Manual...
Страница 141: ...Calibration Reference B CalibrationReference...
Страница 167: ...Calibration Program C CalibrationProgram...
Страница 178: ......
Страница 179: ......