Model 2750 Multimeter/Switch System Service Manual
Calibration
2-21
Mod
el 27
50 Cal
ibr
ation
Unlocking manufacturing calibration
To unlock manufacturing calibration, press and hold in the OPEN key while turning on the
power.
Measuring function generator signal amplitude
The 3Hz function generator signal amplitude must be accurately measured using the digi-
tal multimeter listed in
. Proceed as follows:
1.
Connect the function generator output to the digital multimeter INPUT jacks. (See
for typical connections.)
2.
Turn on the function generator and multimeter, and allow a two-hour warm-up
period before measuring.
3.
Set the function generator to output a 1V RMS sine wave at 3Hz; measure and
record the signal amplitude.
Front panel manufacturing calibration
1.
Install the shorted calibration extender board (see “
”
earlier in this section) in scanner card Slot 1, and select the rear inputs with the
INPUTS switch. Allow three minutes for thermal equilibrium.
2.
Press in and hold the OPEN key while turning on the power.
3.
Press SHIFT then TEST then display CALIB: TEST with the up or down range
key. Press ENTER, select RUN, then enter the appropriate calibration code
(default: 002750).
4.
Select ALL at the CAL:RUN prompt.
5.
Press ENTER to perform the first manufacturing calibration step.
6.
Perform the entire front panel comprehensive calibration procedure discussed ear-
lier in this section. (See “
” earlier in this section.)
7.
Connect the function generator to the Model 2750 front panel INPUT jacks as
shown in
. Select the front input jacks with the INPUTS switch.
Содержание 2750
Страница 13: ...1 PerformanceVerification...
Страница 48: ...1 36 Performance Verification Model 2750 Multimeter Switch System Service Manual...
Страница 49: ...2 Calibration...
Страница 74: ...2 26 Calibration Model 2750 Multimeter Switch System Service Manual...
Страница 75: ...Routine Maintenance 3 RoutineMaintenance...
Страница 86: ...3 12 Routine Maintenance Model 2750 Multimeter Switch System Service Manual...
Страница 87: ...Troubleshooting 4 Troubleshooting...
Страница 111: ...Disassembly 5 Disassembly...
Страница 118: ...5 8 Disassembly Model 2750 Multimeter Switch System Service Manual...
Страница 119: ...Replaceable Parts 6 ReplaceableParts...
Страница 135: ...Specifications A AccuracyandOptimization...
Страница 140: ...A 6 Accuracy and Optimization Model 2750 Multimeter Switch System Service Manual...
Страница 141: ...Calibration Reference B CalibrationReference...
Страница 167: ...Calibration Program C CalibrationProgram...
Страница 178: ......
Страница 179: ......