Model 2750 Multimeter/Switch System Service Manual
Troubleshooting
4-21
Tr
oubleshooting
Table 4-11
4 dry circuit reference signal switching
Range
Q159
Q158
Q157
U133/0.7V
U133/7V
Q123
Q125
1
10
100
1k
ON
ON
ON
ON
ON
ON
ON
ON
ON
ON
ON
ON
OFF
ON
OFF
ON
ON
OFF
ON
OFF
ON
ON
OFF
OFF
ON
ON
OFF
OFF
Range
Q124
Q126
K104*
U133/9
Q171
U208/9
U208/16
1
10
100
1k
OFF
OFF
ON
ON
OFF
OFF
ON
ON
SET
SET
SET
SET
ON
ON
ON
ON
ON
ON
ON
ON
ON
ON
ON
ON
ON
ON
ON
ON
*K104 set states: Pin 8 switched to Pin 7
Pin 3 switched to Pin 4
Table 4-12
DCA signal switching
Range
K103
20mA
100mA
1A
3A
Set
Reset
Reset
Reset
K103 set states:
Pin 8 to 7
Pin 3 to 4
K103 reset states: Pin 8 to 9
Pin 3 to 2
Table 4-13
ACA signal switching
Range
K103
U105
Pin 16
U105
Pin 1
U111
Pin 16
U105
Pin 8
U103
Pin 16
U103
Pin 1
1A
3A
Reset
Reset
ON
ON
ON
ON
OFF
ON
OFF
OFF
OFF
OFF
OFF
OFF
K103 set states:
Pin 8 to 7
Pin 3 to 4
K103 reset states: Pin 8 to 9
Pin 3 to 2
Содержание 2750
Страница 13: ...1 PerformanceVerification...
Страница 48: ...1 36 Performance Verification Model 2750 Multimeter Switch System Service Manual...
Страница 49: ...2 Calibration...
Страница 74: ...2 26 Calibration Model 2750 Multimeter Switch System Service Manual...
Страница 75: ...Routine Maintenance 3 RoutineMaintenance...
Страница 86: ...3 12 Routine Maintenance Model 2750 Multimeter Switch System Service Manual...
Страница 87: ...Troubleshooting 4 Troubleshooting...
Страница 111: ...Disassembly 5 Disassembly...
Страница 118: ...5 8 Disassembly Model 2750 Multimeter Switch System Service Manual...
Страница 119: ...Replaceable Parts 6 ReplaceableParts...
Страница 135: ...Specifications A AccuracyandOptimization...
Страница 140: ...A 6 Accuracy and Optimization Model 2750 Multimeter Switch System Service Manual...
Страница 141: ...Calibration Reference B CalibrationReference...
Страница 167: ...Calibration Program C CalibrationProgram...
Страница 178: ......
Страница 179: ......