Model 2750 Multimeter/Switch System Service Manual
Calibration
2-3
Mod
el 27
50 Cal
ibr
ation
Calibration considerations
When performing the calibration procedures:
•
Make sure that the equipment is properly warmed up and connected to the appro-
priate input jacks. Make sure that the correct front or rear terminals are selected
with the INPUTS switch.
•
Make sure the calibrator is in OPERATE before you complete each calibration
step.
•
Always let the source signal settle before calibrating each point.
•
If an error occurs during calibration, the Model 2750 will generate an appropriate
error message. See
for more information.
WARNING
Observe the following safety precautions when performing these tests:
•
Some of the procedures in this section may expose you to dangerous
voltages. Use standard safety precautions when such dangerous volt-
ages are encountered to avoid personal injury or death caused by elec-
tric shock.
•
For the front panel terminals only, the maximum common-mode volt-
age (voltage between INPUT LO and chassis ground) is 500V peak.
Exceeding this value may cause a breakdown in insulation, creating a
shock hazard.
•
For the plug-in modules, the maximum common-mode voltage (volt-
age between any plug-in module terminal and chassis ground) is listed
in the module’s specifications. Exceeding this value may cause a
breakdown in insulation, creating a shock hazard.
•
When using the front panel terminals simultaneously with plug-in
modules, all cable insulation voltage ratings must equal or exceed the
maximum voltage applied to either the front panel terminals or the
plug-in module terminals.
Содержание 2750
Страница 13: ...1 PerformanceVerification...
Страница 48: ...1 36 Performance Verification Model 2750 Multimeter Switch System Service Manual...
Страница 49: ...2 Calibration...
Страница 74: ...2 26 Calibration Model 2750 Multimeter Switch System Service Manual...
Страница 75: ...Routine Maintenance 3 RoutineMaintenance...
Страница 86: ...3 12 Routine Maintenance Model 2750 Multimeter Switch System Service Manual...
Страница 87: ...Troubleshooting 4 Troubleshooting...
Страница 111: ...Disassembly 5 Disassembly...
Страница 118: ...5 8 Disassembly Model 2750 Multimeter Switch System Service Manual...
Страница 119: ...Replaceable Parts 6 ReplaceableParts...
Страница 135: ...Specifications A AccuracyandOptimization...
Страница 140: ...A 6 Accuracy and Optimization Model 2750 Multimeter Switch System Service Manual...
Страница 141: ...Calibration Reference B CalibrationReference...
Страница 167: ...Calibration Program C CalibrationProgram...
Страница 178: ......
Страница 179: ......