Calibration
2-21
4.
Send the following command to unlock calibration:
:CAL:PROT:CODE
'KI002500'
5.
summarizes the various calibration steps and associated commands,
which are covered in more detail throughout the procedure.
Table 2-7
Remote calibration step summary
Calibrated
function
Calibration command*
Test connections
Input offset
(ch. 1)
Voltage
bias (ch. 1)
Input offset
(ch. 2)
Voltage
bias (ch. 2)
:CAL:PROT:SOUR1:OFFS
:OUTP1 ON
:SOUR1:VOLT:RANG <Range>
:SOUR1:VOLT <+FS>
:CAL:PROT:SOUR1 <DMM_Reading>
:SOUR1:VOLT 0
:CAL:PROT:SOUR1 <DMM_Reading>
:SOUR1:VOLT <-FS>
:CAL:PROT:SOUR1 <DMM_Reading>
:OUTP1 OFF
:OUTP2 ON
:SOUR2:VOLT:RANG <Range>
:SOUR2:VOLT <+FS>
:CAL:PROT:SOUR2:OFFS
:CAL:PROT:SOUR2 <DMM_Reading>
:SOUR2:VOLT 0
:CAL:PROT:SOUR2 <DMM_Reading>
:SOUR2:VOLT <-FS>
:CAL:PROT:SOUR2 <DMM_Reading>
:OUTP2 OFF
Triax shield cap on channel 1 INPUT jack.
DMM to channel 1 OUTPUT/INPUT LO (
DMM to channel 1 OUTPUT/INPUT LO (
DMM to channel 1 OUTPUT/INPUT LO (
DMM to channel 2 OUTPUT/INPUT LO (
Triax shield cap on channel 2 INPUT jack.
DMM to channel 2 OUTPUT/INPUT LO (
DMM to channel 2 OUTPUT/INPUT LO (
Current
(ch. 1)
Current
(ch. 2)
:SENS1:CURR:RANG <Range>
:CAL:PROT:SENS1 <+FS_current>
:CAL:PROT:SENS1 <Zero_current>
:CAL:PROT:SENS1 <-FS_current>
:SENS2:CURR:RANG <Range>
:CAL:PROT:SENS2 <+FS_current>
:CAL:PROT:SENS2 <Zero_current>
:CAL:PROT:SENS2 <-FS_current>
Calibrator to channel 1 INPUT (
)
Calibrator to channel 1 INPUT (
)
Calibrator to channel 1 INPUT (
)
Calibrator to channel 2 INPUT (
)
Calibrator to channel 2 INPUT (
)
Calibrator to channel 2 INPUT (
)
*Procedure repeated separately for each range. Voltage <Range> = 10 or 100. Current <Range> = 2e-9 to 20e-3 in decade steps.
Содержание 2500
Страница 2: ...Contains Calibrating and Servicing Information Model 2500 Dual Photodiode Meter Service Manual...
Страница 11: ......
Страница 13: ......
Страница 15: ......
Страница 16: ...1 PerformanceVerification...
Страница 28: ...2 Calibration...
Страница 56: ...3 RoutineMaintenance...
Страница 59: ...3 4 Routine Maintenance Model 2500 Service Manual...
Страница 60: ...4 Troubleshooting...
Страница 74: ...5 Disassembly...
Страница 81: ...5 8 Disassembly Model 2500 Service Manual...
Страница 82: ...6 ReplaceableParts...
Страница 93: ...6 12 Replaceable Parts Model 2500 Service Manual...
Страница 94: ...A Specifications...
Страница 98: ...B CalibrationReference...
Страница 108: ...C CalibrationProgram...
Страница 117: ......
Страница 119: ......
Страница 120: ......
Страница 121: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...