![Keithley 2500 Скачать руководство пользователя страница 44](http://html1.mh-extra.com/html/keithley/2500/2500_service-manual_3998319044.webp)
Calibration
2-17
Step 6. Channel 2 current calibration
Follow the steps below to calibrate all channel 2 current ranges.
ibration ranges and currents.
1.
Connect the calibrator to the Model 2500 INPUT CHANNEL 2 jack, as shown in
2.
Select the calibrator passive V/R amps function. Set the calibrator output to
+1.90000nA, and turn on the output.
3.
Press the Model 2500 MSR2 key, then set channel 2 to the 2nA current range using
the RANGE
key.
4.
Press the MENU key, select CAL, then press ENTER.
5.
Select EXECUTE, then press ENTER to enter the CAL EXECUTION menu.
6.
Select I-CAL then press ENTER. The unit will display the following:
I -CAL CHANNEL SELECT
CHAN-1 CHAN-2
7.
Select CHAN-2 then press ENTER. The instrument will prompt for the positive
full-range current:
I (+FS)
2
= +2.000000nA
Use
,
,
,
, ENTER or EXIT.
8.
Set the calibrator current to +1.90000nA.
9.
Use the EDIT keys to adjust the Model 2500 display value to agree with the cali-
brator current.
10.
Press ENTER. The unit will prompt for the zero current:
I (0)
2
= +0.000000nA
Use
,
,
,
, ENTER or EXIT.
11.
Set the calibrator current to 0nA then press ENTER. The instrument will prompt
for the negative full-range current:
I(-FS)
2
= -2.000000nA
Use
,
,
,
, ENTER or EXIT.
12.
Set the calibrator current to -1.90000nA.
13.
Adjust the displayed current to agree with the calibrator current, then press the
Model 2500 ENTER key to complete calibration of the present range.
14.
Press EXIT to return to normal display.
15.
Repeat steps 3 through 14 for the 20nA to 20mA ranges, using
guide. Be sure to set the Model 2500 to the correct range using the RANGE
and
keys, and use the corresponding calibrator currents for each range.
Содержание 2500
Страница 2: ...Contains Calibrating and Servicing Information Model 2500 Dual Photodiode Meter Service Manual...
Страница 11: ......
Страница 13: ......
Страница 15: ......
Страница 16: ...1 PerformanceVerification...
Страница 28: ...2 Calibration...
Страница 56: ...3 RoutineMaintenance...
Страница 59: ...3 4 Routine Maintenance Model 2500 Service Manual...
Страница 60: ...4 Troubleshooting...
Страница 74: ...5 Disassembly...
Страница 81: ...5 8 Disassembly Model 2500 Service Manual...
Страница 82: ...6 ReplaceableParts...
Страница 93: ...6 12 Replaceable Parts Model 2500 Service Manual...
Страница 94: ...A Specifications...
Страница 98: ...B CalibrationReference...
Страница 108: ...C CalibrationProgram...
Страница 117: ......
Страница 119: ......
Страница 120: ......
Страница 121: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...