1-8
Performance Verification
7.
Verify that the Model 2500 current reading is within the limits for the 2nA range
shown in
8.
Repeat steps 4 through 7 for the 20nA through 20mA ranges. (Use the calibrator
active amps mode for the 20mA range.)
9.
Repeat steps 4 through 8 for negative input currents of the same magnitude.
10.
After verifying all ranges on channel 1, repeat the entire procedure for channel 2.
Connect the calibrator to the INPUT CHANNEL 2 jack, and select channel 2 by
pressing MSR2. (See
Table 1-2
Current measurement verification limits
Model 2500 range
Calibrator current
Reading limits*
1 year, 65°F to 82°F (18˚C to 28˚C)
2nA
1.90000nA
1.87900 to 1.92100nA
20nA
19.0000nA
18.9220 to 19.0780nA
200nA
190.000nA
189.230 to 190.770nA
2
µ
A
1.90000
µ
A
1.89600 to 1.90400
µ
A
20
µ
A
19.0000
µ
A
18.9610 to 19.0390
µ
A
200
µ
A
190.000
µ
A
189.790 to 190.210
µ
A
2mA
1.90000mA
1.89610 to 1.90390mA
20mA
19.0000mA
18.9790 to 19.0210mA
*Limits do not include calibrator uncertainty.
Содержание 2500
Страница 2: ...Contains Calibrating and Servicing Information Model 2500 Dual Photodiode Meter Service Manual...
Страница 11: ......
Страница 13: ......
Страница 15: ......
Страница 16: ...1 PerformanceVerification...
Страница 28: ...2 Calibration...
Страница 56: ...3 RoutineMaintenance...
Страница 59: ...3 4 Routine Maintenance Model 2500 Service Manual...
Страница 60: ...4 Troubleshooting...
Страница 74: ...5 Disassembly...
Страница 81: ...5 8 Disassembly Model 2500 Service Manual...
Страница 82: ...6 ReplaceableParts...
Страница 93: ...6 12 Replaceable Parts Model 2500 Service Manual...
Страница 94: ...A Specifications...
Страница 98: ...B CalibrationReference...
Страница 108: ...C CalibrationProgram...
Страница 117: ......
Страница 119: ......
Страница 120: ......
Страница 121: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...