2-26
Calibration
1.
Connect the calibrator to the Model 2500 INPUT CHANNEL 1 jack, as shown in
2.
Select the calibrator passive V/R amps function. Set the output current to
1.90000nA, and turn on the calibrator output.
3.
Send the following command to select the channel 1 2nA range:
:SENS1:CURR:RANG 2e-9
4.
Send the following command to calibrate the positive full-range current point:
:CAL:PROT:SENS1 1.9e-9
5.
Set the calibrator current output to 0nA.
6.
Send the following command to calibrate the zero current point:
:CAL:PROT:SENS1 0
7.
Set the calibrator current to -1.90000nA.
8.
Send the following command to calibrate the negative full-range current point:
:CAL:PROT:SENS1 -1.9e-9
9.
Repeat steps 3 through 8 for the 20nA to 20mA ranges using
Be sure to:
•
Select the appropriate range using the :SENS1:CURR:RANG <Range>
command.
•
Send the appropriate calibration current values with the :CAL:PROT:SENS1
<Current> command.
•
Be sure to use the calibrator active amps mode for the 20mA range.
Step 7. Channel 2 current calibration
Follow these steps to calibrate channel 2 current measurements.
summarizes
channel 2 calibration currents and commands.
1.
Connect the calibrator to the Model 2500 INPUT CHANNEL 2 jack, as shown in
2.
Select the calibrator passive V/R amps function. Set the output current to
1.90000nA, and turn on the calibrator output.
3.
Send the following command to select the channel 2 2nA range:
:SENS2:CURR:RANG 2e-9
4.
Send the following command to calibrate the positive full-range current point:
:CAL:PROT:SENS2 1.9e-9
5.
Set the calibrator current to 0nA.
6.
Send the following command to calibrate the zero current point:
:CAL:PROT:SENS2 0
7.
Set the calibrator current to -1.90000nA.
8.
Send the following command to calibrate the negative full-range current point:
:CAL:PROT:SENS2 -1.9e-9
Содержание 2500
Страница 2: ...Contains Calibrating and Servicing Information Model 2500 Dual Photodiode Meter Service Manual...
Страница 11: ......
Страница 13: ......
Страница 15: ......
Страница 16: ...1 PerformanceVerification...
Страница 28: ...2 Calibration...
Страница 56: ...3 RoutineMaintenance...
Страница 59: ...3 4 Routine Maintenance Model 2500 Service Manual...
Страница 60: ...4 Troubleshooting...
Страница 74: ...5 Disassembly...
Страница 81: ...5 8 Disassembly Model 2500 Service Manual...
Страница 82: ...6 ReplaceableParts...
Страница 93: ...6 12 Replaceable Parts Model 2500 Service Manual...
Страница 94: ...A Specifications...
Страница 98: ...B CalibrationReference...
Страница 108: ...C CalibrationProgram...
Страница 117: ......
Страница 119: ......
Страница 120: ......
Страница 121: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...