2-22
Calibration
Step 2. Channel 1 input offset voltage calibration
1.
Install a triax shielding cap on the channel 1 INPUT jack.
2.
Send this command to calibrate channel 1 input offset voltage:
:CAL:PROT:SOUR1:OFFS
Step 3. Channel 1 voltage source calibration
Follow these steps to calibrate the channel 1 voltage bias source.
summarizes
channel 1 calibration voltages and commands.
1.
Connect the DMM to the Model 2500 channel 1 OUTPUT and INPUT LO termi-
nals, as shown in
2.
Select the DMM DC voltage function and enable auto-range.
3.
Send this command to turn on the channel 1 output:
:OUTP1 ON
4.
Send the following command to select the channel 1 10V voltage bias range:
:SOUR1:VOLT:RANG 10
5.
Send the following command to 10V:
:SOUR1:VOLT 10
6.
Note and record the DMM reading, and send that value as the parameter for the fol-
lowing command:
:CAL:PROT:SOUR1 <DMM_Reading>
For example, if the DMM reading is 9.95V, the correct command is:
:CAL:PROT:SOUR1 9.95
7.
Send the following command to output 0V:
:SOUR1:VOLT 0
8.
Note and record the DMM reading, and send that value as the parameter for the fol-
lowing command:
:CAL:PROT:SOUR1 <DMM_Reading>
9.
Send the following command to output -10V:
:SOUR1:VOLT -10
10.
Note and record the DMM reading, and send that value as the parameter for the fol-
lowing command:
:CAL:PROT:SOUR1 <DMM_Reading>
11.
Repeat steps 4 through 10 for the 100V range using
as a guide. Be sure
to:
•
Select the 100V range using the :SOUR1:VOLT:RANG 100 command.
•
Send the ±100V source values where appropriate.
12.
Send this command to turn off the channel 1 output:
:OUTP1 OFF
Содержание 2500
Страница 2: ...Contains Calibrating and Servicing Information Model 2500 Dual Photodiode Meter Service Manual...
Страница 11: ......
Страница 13: ......
Страница 15: ......
Страница 16: ...1 PerformanceVerification...
Страница 28: ...2 Calibration...
Страница 56: ...3 RoutineMaintenance...
Страница 59: ...3 4 Routine Maintenance Model 2500 Service Manual...
Страница 60: ...4 Troubleshooting...
Страница 74: ...5 Disassembly...
Страница 81: ...5 8 Disassembly Model 2500 Service Manual...
Страница 82: ...6 ReplaceableParts...
Страница 93: ...6 12 Replaceable Parts Model 2500 Service Manual...
Страница 94: ...A Specifications...
Страница 98: ...B CalibrationReference...
Страница 108: ...C CalibrationProgram...
Страница 117: ......
Страница 119: ......
Страница 120: ......
Страница 121: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...