![Keithley 2500 Скачать руководство пользователя страница 26](http://html1.mh-extra.com/html/keithley/2500/2500_service-manual_3998319026.webp)
Performance Verification
1-11
Figure 1-3
Connections for channel 1 bias voltage verification tests
8.
Verify the bias voltage output is within the limits for the 10V range shown in
9.
Repeats steps 6 and 8 for an output voltage of -10.000V.
10.
Press the OUTPUT key to turn off the output.
11.
Repeat steps 5 through 10 for the 100V bias voltage range with both +100.00V and
-100.00V output voltages.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS, SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD, REPLACE FUSE WITH SAME TYPE AND RATING.
120
LINE RATING
50, 60Hz
60 VA MAX
FUSE
LINE
630 mAT
(SB)
315 mAT
(SB)
100 VAC
120 VAC
220 VAC
240 VAC
CAT
I
MADE IN
U.S.A.
DIGITAL I/O
RS-232
TRIGGER LINK
INPUT
CHANNEL 1
INPUT
CHANNEL 2
VOLTAGE SOURCE
OUTPUT CHANNEL 1
VOLTAGE SOURCE
OUTPUT CHANNEL 2
RATINGS MAX
100V @ 20mA
COMMON
MODE
200V
RATINGS MAX
100V @ 20mA
!
!
IEEE-488
(CHANGE IEEE ADDRESS
WITH FRONT PANEL MENU)
NEXT
PREV
POWER
DISPLAY
2001
MULTIMETER
DCV
ACV
DCI
ACI
Ω
2
Ω
4
FREQ
TEMP
REL
TRIG
INFO
LOCAL
EXIT
ENTER
RANGE
RANGE
!
F
R
500V
PEAK
FRONT/REAR
2A 250V
AMPS
CAL
STORE RECALL
CHAN
SCAN
FILTER MATH
CONFIG MENU
HI
INPUT
LO
SENSE
Ω
4 WIRE
INPUTS
350V
PEAK
1100V
PEAK
AUTO
Model 2001 DMM
Model 2500
INPUT LO
(Inner Shield)
Triax
Cable
Channel 1
INPUT
Channel 1
OUTPUT
INPUT LO
INPUT HI
Содержание 2500
Страница 2: ...Contains Calibrating and Servicing Information Model 2500 Dual Photodiode Meter Service Manual...
Страница 11: ......
Страница 13: ......
Страница 15: ......
Страница 16: ...1 PerformanceVerification...
Страница 28: ...2 Calibration...
Страница 56: ...3 RoutineMaintenance...
Страница 59: ...3 4 Routine Maintenance Model 2500 Service Manual...
Страница 60: ...4 Troubleshooting...
Страница 74: ...5 Disassembly...
Страница 81: ...5 8 Disassembly Model 2500 Service Manual...
Страница 82: ...6 ReplaceableParts...
Страница 93: ...6 12 Replaceable Parts Model 2500 Service Manual...
Страница 94: ...A Specifications...
Страница 98: ...B CalibrationReference...
Страница 108: ...C CalibrationProgram...
Страница 117: ......
Страница 119: ......
Страница 120: ......
Страница 121: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...