4-wire resistance tests
More precise measurements over a wider range of system and DUT conditions can be
obtained by using the 4-wire measurement scheme shown in
Figure 3-2
. Here, separate sense
leads from the Model 2000 Multimeter are routed through the scanner to the resistor under test.
The extra set of sense leads minimizes the effects of voltage drops across the test leads, greatly
reducing measurement errors with lower DUT resistances. Note, however, that an extra two
poles of switching are required for each resistor tested. For this reason, only five resistors can be
tested using this configuration.
1
DUTs
(5)
R
A. Test Configuration
B. Simplified Equivalent Circuit
5
6
10
1
5
HI
LO
Sense
HI
Sense
LO
HI
Sense HI
LO
Sense LO
2001
2000-SCAN
DUT
2000-SCAN
Ω
Out A
Out B
DMM Rear Panel
(Model 2001/2002 shown)
Figure 3-2
4-wire resistance
test connections
Operation
3-17
Содержание 2000-Scan
Страница 1: ...Model 2000 ScanScanner Card Instruction Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E ...
Страница 12: ......
Страница 13: ...1 General Information General Information ...
Страница 17: ...2 CardConnections andInstallation CardConnec tionsandInstal lation ...
Страница 35: ...3 Operation Operation ...
Страница 58: ...3 24 Operation Section 3 Page 24 Tuesday July 8 2003 2 30 PM ...
Страница 59: ...4 Service Information Service Information ...
Страница 73: ...5 Replaceable Parts Replaceable Parts ...
Страница 76: ...5 4 Replaceable Parts ...
Страница 77: ......
Страница 78: ......
Страница 79: ...A Specifications Specifica tions ...
Страница 84: ......
Страница 85: ......