Output connections to multimeter
After installation, connect the scanner card output leads to the multimeter rear panel jacks as
follows:
WARNING
Signals applied to the front inputs of the multimeter cannot exceed ratings
of the scanner card as long as the scanner card output leads are connected
to the multimeter rear panel input jacks.
•
For 2-pole operation, connect OUT A HI (red) to INPUT HI, and connect OUT A LO
(black) to INPUT LO. See
Figure 2-11
.
NOTE
Be sure to select the rear inputs with the multimeter front panel INPUTS switch when
using the scanner.
Input HI
Out A
HI (Red)
Input LO
Out A
LO (Black)
Note: Card installation in a Model 2001/2002
is shown. For installation in Model 2000,
turn card over.
Figure 2-11
2-pole output
connections
Card Connections and Installation
2-17
Содержание 2000-Scan
Страница 1: ...Model 2000 ScanScanner Card Instruction Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E ...
Страница 12: ......
Страница 13: ...1 General Information General Information ...
Страница 17: ...2 CardConnections andInstallation CardConnec tionsandInstal lation ...
Страница 35: ...3 Operation Operation ...
Страница 58: ...3 24 Operation Section 3 Page 24 Tuesday July 8 2003 2 30 PM ...
Страница 59: ...4 Service Information Service Information ...
Страница 73: ...5 Replaceable Parts Replaceable Parts ...
Страница 76: ...5 4 Replaceable Parts ...
Страница 77: ......
Страница 78: ......
Страница 79: ...A Specifications Specifica tions ...
Страница 84: ......
Страница 85: ......