Output connections
Use the supplied test leads for scanner output connections. Connect red leads to the output
(OUT A and OUT B) HI terminals, and connect black leads to the output LO terminals. See
Figure 2-3
for details. Dress output test leads through the cable clamp, as described in
“Dressing
leads” on page 2-8
. After all wires are connected and secure, close the plastic shield, and secure
it with the locking tab.
NOTE
If you intend to use the scanner card only in the 2-pole mode, it is not necessary to
connect output leads to both OUT A and OUT B. Use only OUT A for the 2-pole mode.
After the scanner card is installed, the output leads must be connected to the multimeter rear
panel input jacks. See
“Output connections to multimeter” on page 2-17
for details.
CH 1
H L
CH 2
H L
CH 3
H L
CH 4
H L
CH 5
H L
OUT A
H L
CH 6
H L
CH 7
H L
CH 8
H L
CH 9
H L
CH 10
H L
OUT B
H L
HI
LO
To DMM
Input
Red
Black
HI
Red
Black
To DMM
Sense
Note
: OUT B connections not required
for 2-pole operation. Plastic shield
not shown.
LO
Figure 2-3
Output connections
Card Connections and Installation
2-7
Содержание 2000-Scan
Страница 1: ...Model 2000 ScanScanner Card Instruction Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E ...
Страница 12: ......
Страница 13: ...1 General Information General Information ...
Страница 17: ...2 CardConnections andInstallation CardConnec tionsandInstal lation ...
Страница 35: ...3 Operation Operation ...
Страница 58: ...3 24 Operation Section 3 Page 24 Tuesday July 8 2003 2 30 PM ...
Страница 59: ...4 Service Information Service Information ...
Страница 73: ...5 Replaceable Parts Replaceable Parts ...
Страница 76: ...5 4 Replaceable Parts ...
Страница 77: ......
Страница 78: ......
Страница 79: ...A Specifications Specifica tions ...
Страница 84: ......
Страница 85: ......