![background image](http://html1.mh-extra.com/html/keithley/2000-scan/2000-scan_instruction-manual_661585024.webp)
Dressing leads
After wires are connected to the terminal blocks, they should be dressed through the cable
clamp as shown in
Figure 2-4
. To do so, unlatch the clip that holds the cable clamp together, then
route all wires flat against the lower half of the clamp. Clamp the wires down, then secure the
clamp with the metal clip unlatched earlier.
CH 1
H L
CH 2
H L
CH 3
H L
CH 4
H L
CH 5
H L
OUT A
H L
CH 6
H L
CH 7
H L
CH 8
H L
CH 9
H L
CH 10
H L
OUT B
H L
Cable Clamp
Note
: Plastic shield not shown.
Metal Clip
Figure 2-4
Routing wires through
cable clamp
2-8