![background image](http://html1.mh-extra.com/html/keithley/2000-scan/2000-scan_instruction-manual_661585066.webp)
Isolation tests
These tests check the leakage resistance (isolation) between all HI and LO terminals and from
HI and LO terminals to chassis ground. In general, the test is performed by applying a voltage
(100V) across the terminals and then measuring the leakage current. The isolation resistance is
then calculated as R = V/I. In the following procedure, the Model 617 functions as a voltage
source and as an ammeter. In the V/I function, the Model 617 internally calculates the resistance
from the known voltage and current levels, and displays the resistance value.
Perform the following steps to check isolation:
1.
Connect a 4” length of Teflon-insulated #22AWG stranded wire to every HI and LO
terminal on the scanner card (both inputs and outputs). Mark the free end of each wire
so you can identify wires when the scanner card is installed.
2.
With the power off, install the scanner card in the multimeter.
3.
Turn on the Model 2001 power.
4.
Turn on the Model 617, and allow the unit to warm up for two hours for rated accuracy.
5.
Select the Model 617 2pA range, and enable zero check and zero correct in that order.
Leave zero correct enabled for the entire procedure.
WARNING
The following steps use high voltage (100V). Be sure to remove power from
the circuit before making connection changes.
6.
Set the Model 617 voltage source for +100V, and select the 200nA current range. Make
sure the voltage source is still in standby.
7.
Place the Model 617 in the V/I measurement function by pressing SHIFT OHMS.
8.
With the Model 617 in standby, connect the electrometer to OUT A HI and LO terminals
of the scanner card, as shown in
Figure 4-3
.
HI
LO
Channel 1
HI
LO
Channel 2-4
Channel 5
HI
LO
Channel 6
HI
LO
Channel 7-9
Channel 10
HI
LO
OUT A
OUT B
HI
LO
2-Pole
4-Pole
Model 2000-SCAN
HI
Model 617
INPUT
Model 7025
Unterminated
Triax Cable
(Red)
Banana to Banana Cable
Ground Link
Removed
Unterminated to
Banana Cable
Source V and
Measure V/I
Com
V-Source LO
V-Source HI
Figure 4-3
Connections for same-
channel isolation tests
4-8
Service Information
Содержание 2000-Scan
Страница 1: ...Model 2000 ScanScanner Card Instruction Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E ...
Страница 12: ......
Страница 13: ...1 General Information General Information ...
Страница 17: ...2 CardConnections andInstallation CardConnec tionsandInstal lation ...
Страница 35: ...3 Operation Operation ...
Страница 58: ...3 24 Operation Section 3 Page 24 Tuesday July 8 2003 2 30 PM ...
Страница 59: ...4 Service Information Service Information ...
Страница 73: ...5 Replaceable Parts Replaceable Parts ...
Страница 76: ...5 4 Replaceable Parts ...
Страница 77: ......
Страница 78: ......
Страница 79: ...A Specifications Specifica tions ...
Страница 84: ......
Страница 85: ......